Sy-B037 Electrophoresis Apparatus Portble Semi-Automatic Medical Device

Classification: Biological Diagnostics
Type: Urine Analyzer
Certification: MSDS
Name: Electrophoresis Analyzer
Current Range: 0~400mA
Voltage Shift: ≤0.2%

Contact Supplier

Diamond Member Since 2018

Suppliers with verified business licenses

Manufacturer/Factory

Basic Info.

Model NO.
SY-B037
Voltage Range
0~600V
Transport Package
Carton
Trademark
Sunnymed
Origin
China
Production Capacity
500 Sets/Month

Product Description

SY-B037 Electrophoresis Apparatus Portble Semi-automatic Medical Device

 
Specification of Electrophoresis Apparatus:
 
Voltage range 0~600V
Current range 0~400mA
Voltage shift ≤0.2%
Current shift ≤0.2%
Working mode  : Constant current or constant voltage
Output format  : in four groups synchronously
Power supply: 220V±10%  50Hz
Fuse: BGXP-5  5A

 

Operating instructions

 

1) Connect the output ends to the gel box. The red end is positive electrode (anode) and the black one is negative electrode (cathode).

 

2) Revolve the knob of "Adjust" anticlockwise until it stops.

 

3) Press the button of "Power" to switch on power supply and lighten the digital panel. 

 

4) Choose a constant mode

 

constant voltage mode

 

   Press the button of "constant T" to turn on the indicator light of "constant T".

 

   Now the electrophoresis apparatus is working on constant voltage mode.

 

constant current mode

 

   Press the button of "constant A" to turn on the indicator light of "constant A".

 

   Now the electrophoresis apparatus is working on constant current mode.

 

5) Revolve the knob of "Adjust" clockwise gently until it reaches a certain value of voltage or current.



 

Cautions

 

1) DO NOT connect the electrophoresis apparatus to the gel box with the power on. Also, DO NOT pull out the output cables that connecting the apparatus to the gel box with the power on. Always Connect the electrophoresis apparatus to the gel box BEFORE turning the power supply ON.

 

2) Assume that you want to set the voltage value to "150V" at the constant voltage mode. If the current value becomes extremely high or even overflows the rated value (400mA) before you could adjust the voltage value to 150V, do examine whether the electrophoresis apparatus is well connected to the gel box or the electrophoresis buffers is accurately prepared. 

 

3) When more than one gel boxes are working together, choose the constant voltage mode to reduce the interactions between boxes.

 

4) PULL the power cables OUT OF the power supply when changing the fuse. Specification of the new fuse should accord with that indicated in "Technical data"


 



Sy-B037 Electrophoresis Apparatus Portble Semi-Automatic Medical Device
Sy-B037 Electrophoresis Apparatus Portble Semi-Automatic Medical Device
Sy-B037 Electrophoresis Apparatus Portble Semi-Automatic Medical Device
Sy-B037 Electrophoresis Apparatus Portble Semi-Automatic Medical Device
Sy-B037 Electrophoresis Apparatus Portble Semi-Automatic Medical Device
Sy-B037 Electrophoresis Apparatus Portble Semi-Automatic Medical Device

Send your message to this supplier

*From:
*To:
avatar Ms. Echo-Sunnymed
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Product Groups

Contact Supplier

Diamond Member Since 2018

Suppliers with verified business licenses

Manufacturer/Factory
Type of Ownership
Limited Company
Management System Certification
ISO 9001, OHSAS/ OHSMS 18001