Customization: | Available |
---|---|
Type: | Infusion Pump |
Certification: | CE |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Pump Type | Drip type infusion pump |
Pumping Way | Peristalic fingers |
Adjustable Drop Rate Range | 1~400 drops/min (step: 1 drop/min) |
Adjustable Volume Rate Range | 1~1200 ml/h (when 1~99.9 ml/h, step: 0.1 ml/h; greater than 99.9 ml/h, step: 1 ml/h) |
Flow Precision | Within±3% (using the IV set appointed by the manufacturer or calibrated high quality IV set) |
Mechanical precision | within ±2%. |
Purge Rate | 800 ml/h |
KVO Rate | 1 ml/h (1~300 ml/h); 3 ml/h (greater than 300 ml/h); |
Infusion Volume Range | 1 ml~9999.9 ml (step: 0.1 ml) |
Maximum Accumulated Volume | 9999.9 ml (step: 0.1 ml) |
Time Range | 1 min~9999 min (step: 1 min) |
Occlusion Alarm Threshold | High: 800 mmHg±200 mmHg (106.7kPa±26.7kPa) Medium: 500 mmHg±100 mmHg (66.7kPa±13.3kPa) Low: 300 mmHg±100 mmHg (40.7kPa±13.3kPa) |
Alarm | Infusion Complete, Empty, Faulty Signal, Misoperation, Occlusion, Door Open, Air Bubble, Low Battery, Setting Error, AC power off, Idle. |
Air Bubble Detector | Method: ultrasonic wave, sensitivity: more than 25 μL |
Fuse | F1AL/250V ("F" indicate fast,"L" indicate low breaking capacity), 2 pcs (installed inside pump). |
Power Supply | AC 85~265V, 50/60Hz; |
Internal Battery | 11.1V rechargeable li-ion battery. |
Capacity | more than 2000mAh; the pump can work more than 4 hours at the flow rate of 25 ml/h after charging for 8 hours (Medium rate specified by GB 9706.27-2005). |
Power | less than 40VA |