Customization: | Available |
---|---|
Type: | Syringe Pump |
Certification: | CE, MSDS |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Features:
1. Applicable syringe: 12 brands for user-defined
2. Occlusion sensitivity: adjustable
3. Can be stacked up and integrated into multi-channel
4. Syringe size: 10, 20, 30, 50 ml
5. Accuracy: ±2%
6. Flow rate: 0.1-1300 ml/h
7. KVO rate: 0.1-1 ml/h in 0.1 ml/h increments, adjustable
Technical Specifications:
Model |
SY-G079-2 Syringe Pump |
Syringe Size |
10, 20, 30, 50 ml |
Applicable Syringe |
2 brands built-in (BD, B. Braun), 10 brands for user-defined |
Volume |
0.1- 9999 ml <1000 ml in 0.1 ml increments ≥1000 mL in 1 ml increments |
Flow Rate |
Syringe 10 ml: 0.1- 400 ml/h Syringe 20 ml: 0.1- 600 ml/h Syringe 30 ml: 0.1- 900 ml/h Syringe 50 ml: 0.1-1300 ml/h <100 ml/h in 0.1 ml/h increments or decrements ≥100 ml/h in 1 ml/h increments or decrements |
Bolus Rate |
Syringe 10 ml: 400 ml/h Syringe 20 ml: 600 ml/h Syringe 30 ml: 900 ml/h Syringe 50 ml: 1300 ml/h |
Accuracy |
±2% |
Infusion Mode |
Flow rate: ml/min, ml/h Time-based Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h |
KVO Rate |
0.1 -1ml/h in 0.1 ml/h increments or decrements, adjustable |
Alarms |
Occlusion, syringe drop off, near empty, end program, low pressure, end battery, AC power off, motor malfunction, unattended pump, syringe moved accidently, pressure sensor error |
Additional Features |
Syringe identification, bolus, real-time shift totals, occlusion pressure release, mute, automatic power switching, residue reminder, system memory |
Occlusion Sensitivity |
High: 0.02-0.06 Mpa Medium: 0.06-0.09 Mpa Low: 0.09-0.12 MPa |
Power Supply, AC |
110/230 V (optional), 50/60 Hz, 20 VA |
Battery |
9.6 ±1.6 V, rechargeable |
Battery Life |
5 hours at 5 ml/h |
Working Temperature |
5- 40°C |
Relative Humidity |
20-90% |
Atmospheric Pressure |
860-1060 hpa |
Size |
314*167*140 mm |
Weight |
2.5 kgs |
Safety Classification |
Class II, type CF |