Customization: | Available |
---|---|
Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
Type: | Anesthesia |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Specification:
Main body | High-strength engineering plastic rack, light,beautiful and corrosion resistant |
Scope of application | adult, child |
Gas source | O2: 0.27~0.55MPa N2O: 0.27~0.55MPa |
Flow meter | O2: 0.1~10L/min N2O: 0.1~10L/min |
O2, N2O linkage and N2O Stopper |
When using nitrous oxide, oxygen concentration>25%; |
Flow rate of rapid oxygen supply | 25~75L/min |
Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. And vaporizers that are imported with original packaging are also available. |
Respiratory circuit | Working mode: all-close, semi-close, semi-open Releasing pressure: 0.5~7 kPa |
Respiratory bellows |
bellows for adults, bellows for children tidal volume range:0~1,500 ml |
Ventilator:
1. Display mode : High-definition 5.7" LCD screen display
2. Ventilation mode:IPPV,SIPPV,SIMV,PEE,MANUAL,SIGH
3. Ventilation parameters:
Tidal volume | 50~1500ml |
Rate | 2~99bmp |
SIMV rate | 2~20bpm |
I:E | 2:1~1:8 |
Inspiratory trigger pressure | -1.0~2.0 kPa |
PEEP | 0~2.0 kPa |
Pressure range | 1.0~6.0 kPa |
SIGH | 1.5 times the inspiratory time |
4. Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate,
SI MV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform,
flow rate - time waveform, PEEP, inspiratory trigger pressure
5. Monitoring of oxygen concentration :(21%~100%)
6. Security Alarm System
Oxygen Concentration alarm |
upper limit setting range (50%~100%) |
Airway pressure alarm |
upper limit setting range 1.0~6.0 kPa low limit setting range:0.4~2.0 kPa |
Per-minute ventilation volume alarm |
upper limit setting range: 3.0~30L/min |
Sustained high-pressure alarm |
it will give alarm when stress have consistently been higher than 2.5 kPa |
Suffocation alarm |
it will give sound and light alarm if there is no tidal volume input for 15 seconds Power alarm |
7. Power: AC 220V±10% 50Hz±1Hz (UPS, Storage battery)